Philips V60 Features
The Philips Respironics V60 Ventilator is an assist ventilator that is intended to augment patient breathing. It is intended for spontaneously breathing individuals who require mechanical ventilation. For example, patients with respiratory failure, chronic respiratory insufficiency, or obstructive sleep apnea in a hospital or other institutional setting under the direction of a physician. The Respironics Philips V60 is intended to support pediatric patients weighing 44 lbs (20 kgs) or greater adult patients.
The Philips V60 BiPAP Ventilator combines Respironics’ ventilation expertise with Philips’ focus on simplifying advanced health care. The result is the noninvasive ventilation gold standard with an invasive ventilation safety net and an interactive display that helps simplify patient management. Because every breath matters and every breath is different, patients need a ventilator that always follows their breathing pattern, whatever their acuity. The Philips Ventilator uses auto-adaptive technology to help ensure patient synchrony and therapy acceptance. The improved signal processing technology is finely tuned for both adult and pediatric patients.
- High resolution graphics facilitating waveform interpretation.
- Internal 6-hour battery enables intra-hospital transport (actual battery time will vary depending on user settings and battery charge level). Standby mode supports patient-clinician interaction without nuisance alarms.
- Data communication capabilities support hospital connectivity.
- Respi-Link, a remote communication tool, allows efficient system diagnostics and upgrades via internet.
- AVAPS maintains a target tidal volume in a pressure limited mode. It provides extra assurance similar to a volume limited mode with the safety of a pressure limited mode.
- CPAP with C-Flex offers three levels of flow-based expiratory pressure relief. This option leads to improved sleep quality and patient comfort, adding greater flexibility and improved treatment acceptance.
- PCV can be used when greater control of the patient’s breathing pattern is required. Additionally, the option of ventilating a patient invasively gives the caregiver greater flexibility in the approach to treatment.
- ICU-grade NIV performance with enhanced safety and monitoring.
- Treatment for a wide range of clinical severities and patients from pediatric to adult.
- Flexibility for intra-hospital transport with an internal 6-hour smart battery.
Philips V60 Specifications
Dimensions
Dimensions listed below with the following configuration: V60 Ventilator with backup battery, ventilator stand, O2 tank holder, circuit arm with mount, patient circuit, O2 manifold with hoses.
- Height: (136 cm)
- Width: (64 cm)
- Weight: (36.1 kg)
Dimensions listed below for only the V60 Unit.
- Height: (33.7 cm)
- Width: (39.4 cm)
- Depth: (42.9 cm)
- Weight (With optional battery): (11.7 kg)
- Weight (Without battery): (10.6 kg)
Patient Types
- Adult
- Pediatric (Less than or equal to 20 kg)
Battery
- 14.4V, 10.0 Ah, 163 Wh
- Maximum system current draw: 11 A
- Charge Voltage: 16.9 V Maximum
- Operating Time: 360 minutes under normal condit
Modes
- Continuous positive airway pressure (CPAP)
- Spontaneous with timed backup (S/T)
- Pressure control ventilation (PCV)
- Average volume assured pressure support (AVAPS) —optional
Settings
- C-Flex OFF, 1–3
- CPAP 4–25 cmH2O
- EPAP 4–25 cmH2O
- IPAP 4–40 cmH2O
- I-time (inspiratory time) 0.30–3.00 sec
- Max P (AVAPS maximum IPAP) 6–40 cmH2O
- Min P (AVAPS minimum IPAP) 5–30 cmH2O
- O2 (oxygen percent) 21–100%
- Ramp time OFF, 5–45 min
- Rate (respiratory rate) 4–60 BPM
- Rise (rise time) 1–5
- Triggering and cycling Auto-adaptive (Auto-Trak)
- AVAPS target tidal volume 200–2000 ml
Monitored Parameters
- Breath Phase / Trigger Indicator Spont, Timed, Exhale
- PIP 0–50 cmH2O
- Patient / Total Leak 0-200 l/min BTPS
- Patient trigger 0–100%
- Respiratory rate 0–90 BPM
- Ti/Ttot 0–91%
- Minute volume ²OPLQ%736
- Tidal volume 0–3000 ml BTPS
General
- Oxygen inlet pressure range 276–600 kPa
- Weight 10.9 kg with optional battery; 10.0 kg without optional battery
- Dimensions 33.7 cm height; 39.4cm weight; 42.9 cm depth
Electrical
- AC voltage 100–240 VAC
- AC frequency 50–60 Hz
- AC power 300 VA
- Battery operating time 6 hours in normal conditions
Environmental Specifications (Operating)
-
- Temperature: (5 to 40 ºC)
Relative Humidity: 15 to 95% (noncondensing) Barometric Pressure: 600 to 765 mmHg (approximately -61 to 1951 m relative to sea level)
Environmental Specifications (Storage)
-
- Temperature: (-20 to 50 ºC)
Relative Humidity: 10 to 95% relative (noncondensing) Barometric Pressure: 600 to 765 mmHg (approximately -61 to 1951 m relative to sea level)
High-pressure Oxygen Supply
-
- Connector: DISS male, DISS female, NIST
- Pressure: 2.76 to 6.00 bar / 276 to 600 kPa / 40 to 87 psig
- Flow: 175 SLPM
- Connector: SIS
- Pressure: 3.31 to 6.00 bar / 331 to 600 kPa / 48 to 87 psig
- Flow: 175 SLPM
High-pressure Oxygen Supply (Using V60 manifold)
-
- Connector: DISS male, DISS female, NIST
- Pressure: 3.10 to 6.00 bar / 310 to 600 kPa / 45 to 87 psig
- Flow: 175 SLPM
- Connector: SIS
- Pressure: 3.66 to 6.00 bar / 366 to 600 kPa / 53 to 87 psig
- Flow: 175 SLPM
Inspiratory Outlet (To patient port)
- Connector: ISO 15 mm female/22 mm male conical
Air Supply
- Integrated blower
Flows and Pressures
- Flow Delivery: 150 L/min at 40 cmH2O at 1951 m altitude (10% degradation in flow at 2286 m
- Flow Range: -240 to 240 L/min BTPS
- Pressure Range: 4 to 40 cmH2O
- Dynamic pressure regulation: ± (2 cmH2O + 4% of target)