Description

Philips Avalon FM20 Features

Avalon FM 20 fetal and maternal monitors are Philips first and only to offer automated coincidence detection (cross-channel verification) using Smart Pulse. It measures the fetal & maternal heart rates separately to enhance diagnostic confidence. The Built-in NST Trace Interpretation, clinical decision support (CDS) application for antepartum monitoring, can help to improve efficiency by sorting out the traces that require closer analysis by the obstetrician. With the Avalon FM Series , you have the confidence of continuous data with a backup memory, battery operation, and LAN interface, combined with smart transducers and a color touchscreen.

  • NST Trace Interpretation to improve efficiency
  • Variety of readings for comprehensive understanding
  • Triplet monitoring option expands capabilities
  • Reliability features allow uninterrupted operation
  • Large, intuitive color display clearly shows status
  • Smart Pulse for automated coincidence detection

Philips Avalon FM20 Specifications

Dimensions

  • M2702A/M2703A
    • Height: 133 mm
    • Width: 286 mm
    • Depth: 335 mm
    • Weight: 5.1 kg
  • M2704A/M2705A
    • Height: 172 mm
    • Width: 420 mm
    • Depth: 370 mm
    • Weight: <9.0 kg

Display

  • Monitor Screen: 16.51 cm
  • Touchscreen Operation: Optional

Battery

  • Operating Time: up to 4 hours

Monitor Operating Environmental Conditions

  • Temperature Range: 0°C to 45°C
  • Humidity Rage: <95% relative humidity @ 40°C
  • Altitude Range:-500 to 3000 m

Monitor Storage Environmental Conditions

  • Temperature Range: -20°C to 60°C
  • Humidity Range: <90% relative humidity @ 60°C
  • Altitude Range: -500 to 13100 m

Transducer Operating Environmental Conditions

  • Temperature Rage: 0°C to 40°C
  • Humidity Range: <95% relative humidity @ 40°C
  • Altitude Range: -500 to 3000 m

Transducer Storage Environmental Contditions

  • Temperature Range: -20°C to 60°C
  • Humidity Range: <90% relative humidity @ 60°C
  • Altitude Range: -500 to 13100 m

SpO2 Environmental Conditions

  • Operating Temperature Range 0°C to 37°C

Fetal / Maternal Ultrasound

  • Measurement Method: Ultrasound Pulsed Doppler
  • Measurement Range: US 50 to 240 bpm
  • Display Resolution: 1 bpm
  • Printer Resolution: 1/4 bpm
  • Jitter @ 200 bpm: ≤ 3 bpm
  • Display Update Rate: 1/ Second
  • Average output power P = (4.3 ± 0.4) mW
  • Peak-negative acoustic pressure p_ = (33.9 ± 3.6) kPa
  • Output beam intensity (Iob) (= spatial average – temporal average intensity) Isata = (2.38 ± 0.75) mW/cm
  • Spatial-peak temporal average intensity I spta = (10.3 ± 2.2) mW/cm2
  • Effective radiating area @ -6 dB 1.81 cm2
  • Beat to Beat Change (max.) for Ultrasound 28 bpm
  • Frequency: 1 MHz ± 100 Hz
  • Signal Range: 3.5 μVpp to 350 μVpp @ 200 Hz

ECG

  • DECG: Single Lead ECG (derived from Fetal Scalp Electrode)
  • MECG: Single Lead ECG (derived from RA and LA electrodes)
  • Measurement Range: 30 to 240 bpm
  • Display Resolution: 1 bpm
  • Recorder Resolution: 1/4 bpm
  • Accuracy: ± 1 bpm or 1%, whichever is greater
  • Beat to Beat Change (max.) 28 bpm
  • Differential Input Impedance > 15MΩ
  • Electrode Offset Potential Tolerance ± 400 mV
  • Filter Bandwidth 0.8 to 80 Hz
  • Inop Auxiliary Current (Leads Off Detection) < 100 μA
  • Input Signal Range DECG: 20 μVpp to 6 mVpp
  • 20 μVpp to 6 mVpp MECG: 150 μVpp to 6 mVpp
  • Dielectric Strength 1500 Vrms
  • Defibrillator Protection None
  • ESU Protection None

Non-invasive blood pressure

  • Measurement Ranges:
    • Systolic 30 to 270 mmHg (4 to 36 kPa)
    • Diastolic 10 to 245 mmHg (1.5 to 32 kPa)
    • Mean 20 to 255 mmHg (2.5 to 34 kPa)
  • Accuracy:
    • Max. Std. Deviation: 8 mmHg (1.1 kPa)
    • Max. Mean Error: ±5 mmHg (±0.7 kPa)
  • Pulse Rate Range Accuracy:
    • 40 to 300 bpm
    • 40 to 100 bpm: ±5 bpm
    • 101 to 200 bpm: ±5% of reading
    • 201 to 300 bpm: ±10% of reading
  • Measurement Time:
    • Typical at HR > 60bpm
    • Auto/manual: 30 seconds (adult)
    • Maximum time: 180 seconds (adult)
  • Cuff Inflation Time Typical for normal adult cuff: Less than 10 seconds
  • Initial Cuff Inflation Pressure 165 ±15 mmHg
  • Auto Mode Repetition Times 1, 2, 2.5, 3, 5, 10, 15, 20, 30, 45, 60 or 120 minutes

Parameters

  • External fetal parameters: US/Toco
  • Twin capability : Standard
  • Triplets capability: Optional
  • Internal fetal parameters: N/A
  • Maternal parameters: MECG, NIBP, Pulse
  • Smart Pulse technology: Standard
  • Cross channel verification: Standard
  • Fetal movement profile: Standard

Readings

  • NST trace interpretation (optional): Up to three Fetal Heart Rates (FHR)

Variety of Readings

  • This antepartum and intrapartum monitor provides a wide range of readings. These include separate maternal pulse measurement; integrated monitoring of maternal pulse rate and blood pressure; external monitoring of fetal heart rates, uterine activity, and fetal movement; and an extensive set of internal fetal parameters such as direct fetal heart rate and uterine pressure. Maternal SpO2 monitoring is included as standard.

Smart Pulse

  • Smart Pulse provides cross-channel verification. It continuously compares all fetal and maternal heart rates to enhance confidence that each is measured separately.

NST Trace Interpretation

  • Built-in NST Trace Interpretation, a clinical decision support (CDS) application for antepartum monitoring, can help to improve efficiency by sorting out the traces that require closer analysis by the obstetrician.

Additional information